MEASUREMENT & CONTROL

pH, FAC & ORP
in HOCl generation.

Compare measurement options and define the monitoring included in your generator proposal.

Read each value
for what it measures.

Keep the measured quantity, reporting basis and units together when comparing results.

pHACIDITY / ALKALINITY

Describes the acidity or alkalinity of the sample.

Read it correctly

pH is not a chlorine concentration or a purity measurement.

FACFREE AVAILABLE CHLORINE

Free available chlorine, measured by a defined method.

Read it correctly

Usually reported as mg/L as Cl₂. Free chlorine and total chlorine are different measurements.

ORPOXIDATION-REDUCTION POTENTIAL

The potential measured by an electrode system, reported in mV.

Read it correctly

ORP depends on the sample chemistry and electrode system. It cannot be converted universally into FAC.

SPECIFY THE MEASUREMENT PACKAGE

What is included
in the monitoring?

Identify the instruments and control functions that support the measurements your team needs.

ArrangementWhat it providesWhat to clarify

Manual sample testing

A result for the sample tested at that time.

Test method, range, consumables and who performs the check.

Online measurement

Readings from an installed measurement system.

Measured parameter, range, sample arrangement and maintenance requirements.

Feedback control

An automatic response to a defined input.

Input signal, control action and the conditions under which it operates.

Match the method
to the sample.

Choose a measurement system for the sample you will test, not simply for the number you want to display.

A method suited to a low-concentration use point may not cover the generated solution directly.

See the measurement locations
01

Measured parameter

Confirm that the method measures free chlorine when FAC is the required result.

02

Working range

Check the range against the expected sample concentration, including relevant operating variation.

03

Sample conditions

Review method limitations and any required sample preparation with the instrument documentation.

04

Routine supplies

Confirm reagents, standards or other consumables, where applicable, and their local availability.

Compare results
on the same basis.

Before comparing two readings, check that the records describe comparable samples and methods.

What a useful measurement record identifies

Sample identification
Where and when the sample was collected.
Method & instrument
The test method, instrument and working range.
Result & reporting basis
The measured quantity, units and any applicable preparation factor.
Relevant conditions
Temperature, pH or other conditions needed to interpret that result.

If results differ, compare these details before concluding that the generator output has changed.

XIAOYA CONTROL FEATURES

Look beyond the display.

Review which operating records and protective functions belong to your proposed configuration.

01:00 · · Measurement to feedback controlDownload video

Operating records

Review the recorded fields available in your proposed interface and how your team will use the operating history and exported data.

Equipment protection

Confirm which alarms and protective shutdown functions are provided for water shortage, feed depletion and a full tank on the selected equipment.

Review testing & validation

A few important distinctions.

Can a pH/ORP display replace a FAC test?

No. Use a method appropriate to free available chlorine when FAC is the required measurement.

Is ORP a universal chlorine-concentration converter?

No. ORP depends on the electrode system and sample chemistry; there is no universal mV-to-FAC conversion.

Does an online reading mean automatic adjustment?

No. Monitoring, alarms, recording and feedback control are separate functions to specify in the proposed package.

Define the monitoring
your project needs.

Start with your application, expected sample range and reporting needs. We can discuss the measurement and control options.

Discuss Monitoring Options